Year
2023
File Attachment
Abstract
This paper presents the results of counting rate and deadtime performance measurements of an
upgraded high efficiency multiplicity counter called the Plutonium Scrap Multiplicity Counter (PSMC)
that uses a high number of compact KM-200 preamplifiers. The measurements were intended to
compare the PSMC’s performance between the original configuration with A-111 preamplifiers and after
its upgrade with KM-200 preamplifiers. The modified system had improved deadtime with lower
deadtime losses. Detector characterization measurements are currently in progress to finalize the
detector operating parameters. The new preamplifiers have the potential to allow measurements in
difficult environments that could extend the range of operation of the neutron multiplicity technique to
spent fuel debris and vitrified waste. The upgraded PSMC also solves many crosstalk and electrical
connectivity issues that limit the old design.