Assay of Low-Level Plutonium Effluents*

Year
1981
Author(s)
S.-T. Hsue - Los Alamos National Laboratory
Faye Hsue - Los Alamos National Laboratory
D.F. Bowersox - Los Alamos National Laboratory
Abstract
In the plutonium recovery section at the Los Alamos National Laboratory, an effluent solution is generated that contains low plutonium concentration and relatively high americium concentration. Nondestructive assay of this solution is demonstrated by measuring the passive L x rays following alpha decay. Preliminary results indicate that an average deviation of 30% between L x-ray and alpha counting can be achieved for plutonium concentrations above 10 mg/L and Am/Pu ratios of up to 3; for plutonium concentrations less than 10 mg/L, the average deviation is 40%. The sensitivity of the L x-ray assay is ~1 mg Pu/L