Publication Date
Volume
17
Issue
3
Start Page
32
File Attachment
V-17_3.pdf4.69 MB
Abstract
The applicability of automatic pipets for dispensing known volumes of water in the volume calibration of accountability tanks of tens-of-liters capacity is demonstrated. Six pipet of capacity in the approximate range 495 to 716 miililiters were calibrated using gravimetric techniques. The relative standard deviation ranged from 0.0015% to 0.0074%. The pipets were used successfully in the volume calibration of accountability tanks of tens-ofliters capacity in the Savannah River Plant.
Additional File(s) in Volume