U-235 Enrichment Determination Via Gamma Spectroscopy Improved Calibration and

Publication Date
Volume
14
Issue
2
Start Page
37
Author(s)
Alan E. Proctor - Idaho National Engineering Laboratory
File Attachment
V-14_2.pdf4.25 MB
Abstract
An improved method of calibrating NDA gamma spectroscopic instrumentation for U-235 enrichment determination is presented. The calibration utilizes a chi-squared minimization method to fit two coefficients to the data. Sources of uncertainty in the initial data are used to "weight" the fit; uncertainties are computed for the two fitting coefficients. The calibration procedure and calculation of "unknown" enrichments is demonstrated. Only simple counting instrumentation is needed; calculations can be readily handled by small computers. An example calculation and program listings are included.
Additional File(s) in Volume
V-14_1.pdf15.64 MB
V-14_2.pdf4.25 MB
V-14_3.pdf4.74 MB