Computer-Determined Assay Time Based On Preset Precision

Year
1994
Author(s)
Lynn A. Foster - Los Alamos National Laboratory
Roland Hagan - Los Alamos National Laboratory
Joseph R. Wachter - Los Alamos National Laboratory
E. Ray Martin - Los Alamos National Laboratory
Abstract
Most current assay systems for special nuclear materials (SNM) operate on the principle of a fixed assay time which provides acceptable measurement precision without sacrificing the required throughput of the instrument. Waste items to be assayed for SNM content can contain a wide range of nuclear material. Counting all items for the same preset assay time results in a wide range of measurement precision and wastes time at the upper end of the calibration range. A short time sample taken at the beginning of the assay could optimize the analysis time on the basis of the required measurement precision. To illustrate the technique of automatically determiningthe assay time, measurements were made with a segmented gamma scanner at the Plutonium Facility of Los Alamos National Laboratory with the assay time for each segment determined by counting statistics in that segment. Segments with very little SNM were quickly determined to be below the lower limit of the measurement range and the measurement was stopped. Segments with significant SNM were optimally assayed to the preset precision. With this method the total assay time for each item is determined by the desired preset precision. This report describes the precision-based algorithm and presents the results of measurements made to test its validity.